Research Article | Volume 24 Issue 1 (, 2018) | Pages 10 - 21
Prosthesis-Specific Predictors of Paravalvular Regurgitation after Transcatheter Aortic Valve Replacement: Impact of Calcification and Sizing on Balloon-Expandable versus Self-Expandable Transcatheter Heart Valves
 ,
 ,
 ,
 ,
 ,
 ,
 ,
 ,
 ,
 ,
 ,
Under a Creative Commons license
PMID : -26182615
Published
Jan. 12, 2015
Abstract

Background and aim of the study: The study aim was to investigate prosthesis-specific predictors for paravalvular aortic regurgitation (PAR) in self-expandable versus balloon-expandable transcatheter heart valves (THVs). PAR is frequently observed after transcatheter aortic valve replacement (TAVR). As self-expandable and balloon-expandable THVs engage differently with the native aortic valve structures, factors that impact PAR may differ between the prosthesis types.

Methods: A total of 137 TAVR patients who underwent pre-procedural multidetector computed tomography and post-procedural transthoracic echocardiography were studied. Predictors for PAR, including annulus area oversizing and aortic valve calcification mass and volume, were analyzed in a multivariate model.

Results: The Medtronic CoreValve (MCV) prosthesis was utilized in 68 patients (50%), and the Edwards SAPIEN (ES) prosthesis in 69 (50%). More than mild PAR was observed in 43 patients (32%). Among MCV patients, aortic valve calcification volume and mass were higher in those with more than mild PAR than in those with none or mild PAR (p = 0.04, p = 0.03, respectively). Among ES patients, the annulus area and perimeter oversizing were lower in those with more than mild PAR compared to those with no or mild PAR (p = 0.001). By univariate and multivariate logistic regression analysis, aortic valve calcification mass was the only independent predictor for PAR in MCV patients (p = 0.02), whereas in ES patients it was THV undersizing (p = 0.002), irrespective of the calcific burden.

Conclusion: For self-expandable THVs, aortic valve calcification mass was the strongest predictor of PAR, whereas for balloon-expandable THVs it was prosthesis undersizing. Hence, in patients evaluated for TAVR these parameters should guide the selection of prosthesis type.

 

 

 

How to cite: aStähli BE, Nguyen-Kim TD, Gebhard C, Erhart L, Frauenfelder T, Tanner FC, Falk V, Landmesser U, Nietlispach F, Lüscher TF, Maier W, Binder RK. Prosthesis-Specific Predictors of Paravalvular Regurgitation after Transcatheter Aortic Valve Replacement: Impact of Calcification and Sizing on Balloon-Expandable versus Self-Expandable Transcatheter Heart Valves. J Heart Valve Dis. 2015 Jan;24(1):10-21. PMID: 26182615.

 
Keywords
Recommended Articles
Research Article
To evaluate the effect of chronic kidney disease on heart rate variability
...
Published: 11/11/2025
Download PDF
Read Article
Research Article
Predictive Accuracy of the Society of Thoracic Surgeons (STS) Score for Postoperative Morbidity and ICU Outcomes after Coronary Artery Bypass Grafting: A Retrospective Study from South India
...
Published: 18/11/2025
Download PDF
Read Article
Research Article
Cardiac Manifestations in Systemic Lupus Erythematosus: A Prospective Observational Study from A Tertiary Care Centre
...
Published: 18/11/2025
Download PDF
Read Article
Research Article
To Compare improvement in Low vision Quality of Life (LVQOL) in children, having low vision due to myopia, with optical aids versus non optical aids
Published: 29/10/2025
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease