High Transcatheter Valve Replacement May Reduce Washout in the Aortic Sinuses: an In-Vitro Study
 ,
 ,
 ,
Under a Creative Commons license
PMID : -26182616
Published
Jan. 12, 2015
Abstract

Background and aim of the study: Transcatheter aortic valve replacements (TAVRs) are performed using fluoroscopic guidance, which makes precise positioning challenging. The aim of the present study was to investigate the effect of TAV positioning on flow characteristics in the ascending aorta and sinuses.

Methods: A commonly used TAV design with a supra-annular support section was investigated using particle image velocimetry (PIV) under physiological flow and pressure conditions. A bioprosthetic valve served as a control and mimicked the native aortic valve. The valve assembly was mounted in a custom-designed chamber with an axisymmetric sinus, the design of which was based on anatomic dimensions. The TAV was deployed in the supra-annular (high) and sub- annular (low) implantation positions and studied at two cardiac outputs (5.0 and 2.5 l/min).

Results: The TAV showed good systolic flow characteristics with wide forward flow jets in the ascending aorta (V = 1.5 m/s at 5.0 1/min; V = 1.0 m/s at 2.5 1/min). In the high implantation, the physical spacing between the leaflet free edge and sinotubular junction was reduced (< 10 mm), causing a weaker sinus vortex and a lower washout. A larger region of low velocity (< 0.1 m/s) in the sinus was observed at high implantation at all time points, particularly at a cardiac output of 2.5 l/min. In the low implantation, good sinus washout was observed.

Conclusion: For optimal sinus washout, sub- annular TAV deployment is recommended, particularly for patients with smaller sinuses. An impaired cardiac output may also require sub- annular deployment for adequate sinus washout. The study results confirmed the need for precise TAV deployment tailored to patient-specific annular and sinus dimensions, as sub-optimal positioning may inhibit coronary perfusion and cause potential regions of stasis near the aortic annulus.

 

 

 

How to cite: Kumar G, Raghav V, Lerakis S, Yoganathan AP. High Transcatheter Valve Replacement May Reduce Washout in the Aortic Sinuses: an In-Vitro Study. J Heart Valve Dis. 2015 Jan;24(1):22-9. PMID: 26182616.

 
Keywords
Recommended Articles
Research Article
Conch Blowing (Shankhnaad) As an Adjunct Intervention for Speech Fluency Disorders: A Randomized Controlled Trial
Published: 20/10/2025
Download PDF
Read Article
Research Article
Clinico-radiological evaluation and management of lower extremity deep vein thrombosis in a tertiary care centre; a retrospective observational study
...
Published: 20/10/2025
Download PDF
Read Article
Research Article
A Prospective Study on Heart Rate Variability as an Early Marker of Metabolic Syndrome in Young Adults
Published: 31/12/2024
Download PDF
Read Article
Research Article
Depression and Anxiety as Predictors of Major Adverse Cardiac Events in Patients with Coronary Artery Disease
Published: 30/12/2023
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease