Research Article | Volume 5 Issue 1 (, 1999) | Pages 66 - 73
Comparison of magnetic resonance imaging and Laser Doppler Anemometry velocity measurements downstream of replacement heart valves: implications for in vivo assessment of prosthetic valve function
 ,
 ,
 ,
 ,
 ,
 ,
 ,
1
Cardiovascular Fluid Mechanics Laboratory, Georgia Institute of Technology, Altanta, GA, USA.
Under a Creative Commons license
PMID : -8834728
Published
Jan. 22, 1996
Abstract

Background and aim of the study: The non-invasive, in-vivo assessment of prosthetic valve function is compromised by the lack of accurate measurements of the transvalvular flow fields or hemodynamics by current techniques. Short echo time magnetic resonance imaging (MRI) may provide a method for the non-invasive, in vivo assessment of prosthetic valve function by accurately measuring changes in the transvalvular flow fields associated with normal and dysfunctional prosthetic valves. The objectives of these in vitro experiments were to investigate the potential for using MRI as a tool to measure the complex flow fields distal to replacement heart valves, and to assess the accuracy of MRI velocity measurements by comparison with Laser Doppler Anemometry (LDA), a gold standard.

Methods: The velocity fields downstream of tilting disc, bileaflet, ball and cage, and pericardial tissue valves were measured using both three-component LDA and MRI phase velocity encoding under a steady flow rate of 22.8 l/min, simulating peak systolic flow. The valves were tested under normal and stenotic conditions to assess the MRI capabilities under a wide range of local flow conditions, velocities and turbulence levels. A new short echo time MRI technique (FAcE), which allowed velocity measurements in stenotic jets with high turbulence, was tested.

Results: Good overall agreement was obtained between the MRI velocity measurements and the LDA data. The MRI velocity measurements adequately reproduced the spatial structure of the flow fields. In most cases peak velocities were accurately measured to within 15%.

Conclusions: The results indicate that the FAcE MRI method has the potential to be used as a diagnostic tool to assess prosthetic valve function.

 

 

 

How to cite: Fontaine, A. A., Heinrich, R. S., Walker, P. G., Pedersen, E. M., Scheidegger, M. B., Boesiger, P., Walton, S. P., & Yoganathan, A. P. (1996). Comparison of magnetic resonance imaging and Laser Doppler Anemometry velocity measurements downstream of replacement heart valves: implications for in vivo assessment of prosthetic valve function. The Journal of heart valve disease5(1), 66–73.

 
Keywords
Recommended Articles
Research Article
A Prospective Hospital-Based Cross-Sectional Observational Study on Sonographic Assessment of Fetal Kidney Length as a Predictor of Gestational Age in the Second and Third Trimester at a Tertiary Care Centre in Indore
...
Published: 28/10/2025
Download PDF
Read Article
Research Article
To study Cholecystectomy in esource limited settings
...
Published: 28/10/2025
Download PDF
Read Article
Research Article
Changes in liver function tests and incidence of acute liver injury after on-pump CABG surgery
...
Published: 28/10/2025
Download PDF
Read Article
Research Article
A Study to Compare Maternal and Fetal Outcomes in Antenatal Women Undergoing Elective and Emergency Caesarean Section at A Tertiary Care Center in North West Rajasthan
...
Published: 12/10/2025
Download PDF
Read Article
© Copyright Journal of Heart Valve Disease